FDA Recall Terminated

TRU-Balance 3 Power Positioning Systems, TB3 Lift/Tilt/Recline (Gen 1). Mobility Medical Device

Recall: Z-0942-2017 · Initiated December 1, 2016

Recall

Recall Number
Z-0942-2017
Event Number
75881
Firm
Pride Mobility Products Corp
FEI Number
3005068474
Product Code
ITI
Status
Terminated
Root Cause
Use error
Initiated
December 1, 2016
Posted
January 3, 2017
Terminated
March 19, 2020
Address
182 Susquehanna Ave, Exeter, PA, 18643-2653

Description

TRU-Balance 3 Power Positioning Systems, TB3 Lift/Tilt/Recline (Gen 1). Mobility Medical Device

Reason

The interface between the TB3 back and tilt base can separate when tilted beyond the specified angle.

Action

Pride Mobility Products Corp. mailed a letter to affected customers to inform them of the issue. Systems were not required to be returned. Customers were informed that a TB3 Seat Depth Adjustment Reinformcement Kit which contains (2) reinforcement plates and associated hardware to strengthen the interface will be installed to address the issue. Direct accounts were asked to contact patients to inform them of the product correction. Customers were also requested to complete and return Pride's completion card once the correction has been made. Pride will issue a $65.00 credit once the completion card is received. Customers with questions were instructed to call Pride's Upgrade Department at 866-222-6212. For questions regarding this recall call 800-800-8586.

Distribution

Nationwide Distribution

Quantity

12,343 chairs