The WinStation Retinal Imager, also known as Merge Eye Station. DCF-1024 & DCF-512 Digital Imaging System Product Usage: The WinStation Retinal Imager has the same intended use as other fundus cameras. It is specifically used to produce color or black and white images of the retina (fundus imaging) and anterior segement (slit lamp imaging), and fluorescein and ICG angiographic images in a non-invasive manner.
Recall
- Recall Number
- Z-0939-2019
- Event Number
- 81869
- Firm
- Merge Healthcare, Inc.
- FEI Number
- 1000116130
- Product Code
- HKI
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- December 20, 2018
- Posted
- February 25, 2019
- Terminated
- February 28, 2023
- Address
- 900 Walnut Ridge Dr, Hartland, WI, 53029-8347
Description
The WinStation Retinal Imager, also known as Merge Eye Station. DCF-1024 & DCF-512 Digital Imaging System Product Usage: The WinStation Retinal Imager has the same intended use as other fundus cameras. It is specifically used to produce color or black and white images of the retina (fundus imaging) and anterior segement (slit lamp imaging), and fluorescein and ICG angiographic images in a non-invasive manner.
Certain configurations of Merge Eye Station hardware may result in electrical output that exceeds the allowable limits for medical devices. Leakage current was found to exceed allowable limits in earth leakage current, touch current in a single fault condition (of open ground) and patient leakage current in a single fault condition (of open ground).
On about 12/20/2018 Merge Healthcare notified customers via "Urgent Field Safety Notice." The notice informed customers that certain configurations of Merge Eye Station hardware may result in electrical output that exceeds the allowable limits for medical devices. A leakage current was found to exceed allowable limits in earth leakage current, touch current in a single fault condition (of open ground) and patient leakage current in a single fault condition (of open ground). Instructions included to immediately cease use of all Eye Station hardware and that a Merge Healthcare representative will contact each customer to schedule a correction once the modification is available. Questions regarding the recall notification can be directed to customer support at [email protected] or 877-741-5369.
Distribution nationwide to Alabama, Alaska, Arizona, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, , Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West , Virginia, Wisconsin. International distribution to Algeria, Austria, Belgium, Canada, Czech Republic, Denmark, Germany, Israel, Netherlands, Puerto Rico, United Kingdom.
1,252 Units