Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system.

Recall: Z-0939-2013 · Initiated December 13, 2012

Recall

Recall Number
Z-0939-2013
Event Number
64447
Firm
Draeger Medical, Inc.
FEI Number
2517967
Product Code
MHX
Status
Terminated
Root Cause
Labeling design
Initiated
December 13, 2012
Posted
March 11, 2013
Terminated
May 20, 2013
Address
3135 Quarry Rd, Telford, PA, 18969-1042

Description

Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system.

Reason

The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a non-vertical position, a vertical position but upside down, or laying it on the floor or on the patient bed.

Action

Drager sent an "URGENT MEDICAL DEVICE RECALL" letter dated December 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 1-800-543-5047 for questions relating to this notice.

Distribution

Distributed in the states of CA, IL, and MA.

Quantity

260