therascreen PIK3CA RGQ PCR Kit (23) US IVD product- aid clinicians in identifying breast cancer patients who may be eligible for treatment with PIQRAY (alpelisib) based on a PIK3CA Mutation Detected result. REF 873121
Recall
- Recall Number
- Z-0937-2021
- Event Number
- 87047
- Firm
- Qiagen Sciences LLC
- FEI Number
- 3003572099
- Product Code
- OWD
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- December 11, 2020
- Terminated
- April 22, 2022
- Address
- 19300 Germantown Rd, Germantown, MD, 20874-1415
Description
therascreen PIK3CA RGQ PCR Kit (23) US IVD product- aid clinicians in identifying breast cancer patients who may be eligible for treatment with PIQRAY (alpelisib) based on a PIK3CA Mutation Detected result. REF 873121
Kit may generate false Q546R mutation positive results caused by nonspecific molecular interactions within the Q546R higher than previously observed and described within the Instructions For Use.
Qiagen US notified customers by Urgent Medical Device Correction letter via e-mail on 12/11/20. The e-mail additionally contains the letter as PDF. Letter states reason for recall, health risk and action to take: For patient samples where a PIK3CA Mutation Detected Result is obtained, disregard Q546R positive results. Cease to report patient samples where a Q546R mutation result is obtained as PIK3CA Mutation Detected. Information about individual target results is specifically obtained from the Individual target result column of the Rotor-Gene Assay Manager v2.1 result table. If multiple mutations including Q546R are detected, disregard the Q546R result only. Continue to consider all other results as valid and report them accordingly. l For the purpose of run validity, Q546R control data must still be produced. The therascreen PIK3CA RGQ PCR Kit software requires valid run control data from the Q546R reaction for the overall test to be valid. Therefore, continue to use all supplied reagents and perform testing as described in the kit handbook, but disregard Q546R positive results generated for patient samples as described in the previous bullet points. l Forward this information to all individuals and departments within your organization who are using the therascreen PIK3CA RGQ PCR Kit REF 873121. If you are not the end-user, please forward this notice to the product end-user. l Review this notice with your laboratory/medical director. l Complete the Acknowledgement or Receipt Form attached to this letter by December 22, 2020 and email it to [email protected]. QIAGEN is now revising the Instructions for Use for the therascreen PIK3CA RGQ PCR Kit, as described in this notice, to reduce any risk resulting from non-specific molecular interactions within the Q546R reaction leading to the generation of Q546R false mutation positive results. QIAGEN is also updating the therascreen PIK3CA FFPE and Plasma Assay Profile software packages to
CA, IN, MI, MN, NC, NJ, NM, OH, TX
575 kits US