FDA Recall
Terminated
Churchill Medical Systems, Inc. A Vygon Company - All sizes of CMS-CT6T225 6F Triple Lumen CT PICC with M.S.T. Components, Peripherally Inserted Central Catheter.
Recall: Z-0937-2013
·
Initiated December 17, 2012
Recall
- Recall Number
- Z-0937-2013
- Event Number
- 64458
- Firm
- Vygon Corporation
- FEI Number
- 2245270
- Product Code
- LJS
- Status
- Terminated
- Root Cause
- Component change control
- Initiated
- December 17, 2012
- Posted
- March 11, 2013
- Terminated
- January 14, 2014
- Address
- 2750 Morris Rd, Ste A200, Lansdale, PA, 19446-6083
Description
Churchill Medical Systems, Inc. A Vygon Company - All sizes of CMS-CT6T225 6F Triple Lumen CT PICC with M.S.T. Components, Peripherally Inserted Central Catheter.
Reason
Product was packaged with the incorrect introducer needle. The kit label indicates a "safety" needle is included; the kits were packaged with a non-safety needle.
Action
Churchill Medical Systems sent a letter to their consignees on December 19, 2012.
Distribution
Product was distributed to 4 hospitals (under one hospital network) within the State of Pennsylvania
Quantity
210