FDA Recall
Terminated
Potassium Electrode Tip; Part No.s: 668281 Tip Assembly- Synchron Family 669117 Packaged- Synchron Family 441104 Electrode Assembly- EL-ISE 445604 Electrode Packaged- EL-ISE
Recall: Z-0937-05
·
Initiated May 20, 2005
Recall
- Recall Number
- Z-0937-05
- Event Number
- 32164
- Firm
- Beckman Coulter Inc
- FEI Number
- 2050012
- Product Code
- CEM
- Status
- Terminated
- Root Cause
- Other
- Initiated
- May 20, 2005
- Posted
- June 28, 2005
- Terminated
- November 21, 2005
- Address
- 200 S Kraemer Blvd, Brea, CA, 92822-6208
Description
Potassium Electrode Tip; Part No.s: 668281 Tip Assembly- Synchron Family 669117 Packaged- Synchron Family 441104 Electrode Assembly- EL-ISE 445604 Electrode Packaged- EL-ISE
Reason
Does not meet performance criteria and exhibits upward trend of result recoveries during calibration period.
Action
Firm sent letters by US Mail requesting that users discard thier probes.
Distribution
United States and Canada
Quantity
2686