FDA Recall Terminated

Boston Scientific, Impulse Angiographic Catheter, Sterile EO. Catheters are single lumen catheters with a single proximal hub for attachment to a dye delivery system. Catalog No. 16599-89. 6F AR MOD (Single Pack). Distal curve AR MOD. 100 cm length. Made in Mexico: Parque industrial Le Mesa, Tijuana, Baja California 22050 MEX. Boston Scientific Corporation, Natick, MA 01760-1537 USA.

Recall: Z-0936-2008 · Initiated November 20, 2007

Recall

Recall Number
Z-0936-2008
Event Number
45903
Firm
Boston Scientific
FEI Number
3002095335
Product Code
DQO
Status
Terminated
Root Cause
Release of Material/Component prior to receiving test results
Initiated
November 20, 2007
Posted
January 31, 2008
Terminated
October 23, 2008
Address
1 Scimed Pl, Maple Grove, MN, 55311-1565

Description

Boston Scientific, Impulse Angiographic Catheter, Sterile EO. Catheters are single lumen catheters with a single proximal hub for attachment to a dye delivery system. Catalog No. 16599-89. 6F AR MOD (Single Pack). Distal curve AR MOD. 100 cm length. Made in Mexico: Parque industrial Le Mesa, Tijuana, Baja California 22050 MEX. Boston Scientific Corporation, Natick, MA 01760-1537 USA.

Reason

Flash may be protruding from the lumen of the catheter shaft. If this flash is present in the catheter, there is the potential of an embolization resulting from the flash detaching from the device during the procedure.

Action

The firm sent an Urgent Medical Device Recall letter dated 11/20/2007, to all affected hospitals. The letter describes the defect with the catheters and Product Table information including UPNs, Catalog Numbers, Lot/Batch Numbers and Product Descriptions. The firm requests that further distribution or use of any remaining product affected by the recall should cease immediately. ****IMPORTANT INFORMATION: If it is a practice of the hospital to remove product from the outer carton and store on shelves in the inner-pouch only. Users are to review the Product Table carefully and consider both the inner and outer packaging product codes when search for affected/recalled product.****

Distribution

Nationwide Distribution including states of AL, CA, FL, GA, KS, KY, LA, MN, NV, OH, OK, PA, SD, TN, TX, and WI.

Quantity

211 total single units