FDA Recall Terminated

Abbott Latex-Free Primary I.V. Plumset Convertible Pin, 95 inch with Injection Site and Secure Lock, No. 19020; Made in Costa Rica, Abbott Laboratories, North Chicago, IL 60064

Recall: Z-0936-03 · Initiated May 13, 2003

Recall

Recall Number
Z-0936-03
Event Number
26349
Firm
Abbott Laboratories HPD/ADD
FEI Number
1415939
Product Code
FPA
Status
Terminated
Root Cause
Other
Initiated
May 13, 2003
Posted
June 17, 2003
Terminated
November 20, 2003
Address
100/200 Abbott Park Road, Abbott Park, IL, 60064

Description

Abbott Latex-Free Primary I.V. Plumset Convertible Pin, 95 inch with Injection Site and Secure Lock, No. 19020; Made in Costa Rica, Abbott Laboratories, North Chicago, IL 60064

Reason

The I.V. administration set incorrectly lists the priming volume from the cassette inlet to distal male adapter as 13.5 mL instead of 11.3 mL

Action

Recall letter dated 5/12/03 sent to the account on 5/13/03, informing them of the labeling discrepancy and requesting the return of any inventory of the affected lot to Abbott for credit.

Distribution

Miami, FL

Quantity

4,656 units