FDA Recall Terminated

Siemens Total Iron Binding Capacity (IBCT) Flex Reagent Cartridge. An in vitro diagnostic test intended to quantitatively measure Total Iron Binding Capacity in human serum.

Recall: Z-0933-2013 · Initiated December 20, 2012

Recall

Recall Number
Z-0933-2013
Event Number
64063
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2517506
Product Code
JMO
Status
Terminated
Root Cause
Process design
Initiated
December 20, 2012
Posted
March 11, 2013
Terminated
August 26, 2014
Address
500 Gbc Dr PO BOX 6101, Ms 514, Newark, DE, 19702-2466

Description

Siemens Total Iron Binding Capacity (IBCT) Flex Reagent Cartridge. An in vitro diagnostic test intended to quantitatively measure Total Iron Binding Capacity in human serum.

Reason

The analytical sensitivity claim of 6 ug/dL (1.1 umol/L) in the Instructions for Use (IFU) is no longer valid. Siemens is conducting a voluntary corrective action for Dimension Total Iron Binding Capacity (IBCT) to modify the analytical sensitivity claim to 36 ug/dL (6.44 umol/L).

Action

Siemens Healthcare sent a "Urgent Medical Device Correction" letter dated Dec 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were notified that theanalytical sensitivity of 6 ug/dL (1.1 umol/L) in the Dimension(R) IBCT FLex(R) Reagent Cartridge instructions for us (IFU) is no longer valid. Please make adjustments to middle wave products or Laboratory Information Systems ( LIS ) as appropriate for your Laboratory. Please complete the attached form and fax it to (302) 631-8467, to indicate you have received this information. If you have any questions please contact your Siemens Technical Solutions Center or your local Siemens technicial support representative at (302) 631-6311.

Distribution

Nationwide Distribution

Quantity

19, 503 cartons