Linemaster Wireless Footswitch Accessory, Model SP-9970214-005112010 and Model SP-9970214-005112232. Sound Surgical Technologies, LLC, Louisville, CO 80027.

Recall: Z-0933-2008 · Initiated December 28, 2007

Recall

Recall Number
Z-0933-2008
Event Number
46293
Firm
Sound Surgical Technologies, LLC
FEI Number
3004106598
Product Code
GEI
Status
Terminated
Root Cause
Device Design
Initiated
December 28, 2007
Posted
March 13, 2008
Terminated
April 1, 2008
Address
357 Mccaslin Blvd, Ste 100, Louisville, CO, 80027-2951

Description

Linemaster Wireless Footswitch Accessory, Model SP-9970214-005112010 and Model SP-9970214-005112232. Sound Surgical Technologies, LLC, Louisville, CO 80027.

Reason

Failure to shut-off: Surgical equipment, activated by a wireless foot switch with VASER system, may remain powered on when switch is no longer depressed.

Action

Consignees were sent an Advisory Notice on 12/28/2007. The notification describes the defect. Sound Surgical will be contacting facilities to arrange a service call to install new cables between the footswitch receiver and the VASER system. Facilities may contact the Technical Service Center at 1-888-677-8765.

Distribution

Worldwide Distribution including USA states of AL, CA, CT, FL, GA, IN, KY, LA, MI, NJ, NY, OH, OK, PA, TX, and VA. Foreign Distribution to Colombia, Costa Rica and Switzerland.

Quantity

127 switches