Bayer Contour TS blood glucose test strips packaged in packs of 10, 25, 50 and 100 strips; Bayer Healthcare, Mishawaka, IN. U.S. Product codes 1820, 1823, 1825 and 9578; 1835 and 1836 (Korea); 1832, 1833 and 1837 (India); 1823 and 1825 (Mexico).
Recall
- Recall Number
- Z-0932-2008
- Event Number
- 46207
- Firm
- Bayer Healthcare LLC
- FEI Number
- 1826988
- Product Code
- LFR
- Status
- Terminated
- Root Cause
- Process design
- Initiated
- December 21, 2007
- Posted
- March 13, 2008
- Terminated
- July 7, 2009
- Address
- 430 S Beiger Street, Mishawaka, IN, 46544-3207
Description
Bayer Contour TS blood glucose test strips packaged in packs of 10, 25, 50 and 100 strips; Bayer Healthcare, Mishawaka, IN. U.S. Product codes 1820, 1823, 1825 and 9578; 1835 and 1836 (Korea); 1832, 1833 and 1837 (India); 1823 and 1825 (Mexico).
Inaccurate Test Strips Results: Results in blood glucose readings with a positive bias are outside the product specifications. Patient test results may demonstrative results 5 percent to 17 percent higher.
The firm issued a Press Release on 12/21/07. A Test Strip Recall letter was also sent to the firm's consignees and health care providers instructing them that use of the product should cease and that health care providers contact the firm for replacements. Additional information can be found at www.bayerdiabeties.com or by contacting the Bayer at 1-800-348-8100.
Worldwide Distribution --- including USA and countries of Austria, France, India, Korea, Mexico, Philippines and Turkey.
97,000 bottles of test strips