FDA Recall Terminated

ACUSON S2000 Automated Breast Volume Scanner (ABVS); Model number 10042541 Distributed by Siemens Medical Solutions USA Inc. business Unit: Ultrasound. The S2000 Automated Breast Volume Scanner (ABVS) Ultrasound System for imaging of the breast allows volumetric acquisition of the breast image.

Recall: Z-0930-2013 · Initiated January 18, 2013

Recall

Recall Number
Z-0930-2013
Event Number
64231
Firm
Siemens Medical Solutions USA, Inc.
FEI Number
2936884
Product Code
IYO
Status
Terminated
Root Cause
Process design
Initiated
January 18, 2013
Posted
March 9, 2013
Terminated
July 16, 2013
Address
685 E Middlefield Rd, Mountain View, CA, 94043-4045

Description

ACUSON S2000 Automated Breast Volume Scanner (ABVS); Model number 10042541 Distributed by Siemens Medical Solutions USA Inc. business Unit: Ultrasound. The S2000 Automated Breast Volume Scanner (ABVS) Ultrasound System for imaging of the breast allows volumetric acquisition of the breast image.

Reason

On some ACUSON S2000 ABVS systems, the inside of the transducer pod may contain sharp edges.

Action

Siemens sent a Customer Safety Advisory notification letter to Siemens Service Personnel on January 18, 2013 and which would be provided to all affected sites during their visit to correct the problem. The notification identified the product, the problem and the action to be taken. The local Customer Service Engineer will perform the test at their site. If the system passes the test, no further action is required. If the ABVS pod fails the inspection, the Service Engineer will correct the sharp edge. For questions regarding this recall call 650-694-5398.

Distribution

Worldwide Distribution - USA including WV and OH; Internationally to Australia, Austria, Bahrain, Belgium, Brazil, China (P.R.) Estonia, France, Germany, Greece, India, Indonesia, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Macedonia, Malaysia, Mexico, Netherlands, Peru, Philippines, Poland, Portugal, Qatar, Rep of Korea, Romania, Sweden, Taiwan, Thailand, Turkey, Unite Arab Emirates, United Kingdom and Venezuela.

Quantity

214 affected devices