FDA Recall Open, Classified

BrightView, gamma camera system; Model Nos.: 6-digit format 882480; 12-digit format 453560279781 453560279791 453560279811 453560279801; 4x4 digit format 2170-3000A 2170-3001A 2170-3002A 2170-3003A

Recall: Z-0929-2024 · Initiated December 15, 2023

Recall

Recall Number
Z-0929-2024
Event Number
93767
Firm
Philips North America
FEI Number
3006648320
Product Code
KPS
Status
Open, Classified
Root Cause
Component design/selection
Initiated
December 15, 2023
Posted
February 2, 2024
Address
222 Jacobs St, Cambridge, MA, 02141-2296

Description

BrightView, gamma camera system; Model Nos.: 6-digit format 882480; 12-digit format 453560279781 453560279791 453560279811 453560279801; 4x4 digit format 2170-3000A 2170-3001A 2170-3002A 2170-3003A

Reason

Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.

Action

On December 15, 2023, the firm notified affected customers via URGENT Medical Device Correction letters. Customers were instructed to not position a patient's lower limbs directly under the detector below the Center of Gantry bore. Customers should circulate the notice to all users of the device for awareness and retain the letter with affected systems until the correction is complete. Philips will contact customers to schedule a time for a Philips Field Service Engineer (FSE) to visit the customer site and correct the system if necessary. If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377). As of 3/14/24, Philips will begin inspecting all affected devices in the field to ensure alignment of the radial leadscrew. Systems that are identified as having a misaligned leadscrew will undergo a report. Additionally, Philips will begin intsallation of a secondary catching mechanism (safety plate) starting in May 2024 in all devices. The firm will contact customers to coordinate these activities.

Distribution

Domestic distribution nationwide. International distribution worldwide.

Quantity

302 US; 264 OUS