Medtronic Cobalt DR MRI SureScan DDPB3D1, MR Conditional implantable dual chamber cardioverter defibrillator with SureScan technology and Bluetooth wireless telemetry (DDE-DDDR)
Recall
- Recall Number
- Z-0928-2021
- Event Number
- 86887
- Firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- FEI Number
- 2182208
- Product Code
- NIK
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- November 15, 2020
- Terminated
- December 23, 2025
- Address
- 8200 Coral Sea St Ne, Mounds View, MN, 55112-4391
Description
Medtronic Cobalt DR MRI SureScan DDPB3D1, MR Conditional implantable dual chamber cardioverter defibrillator with SureScan technology and Bluetooth wireless telemetry (DDE-DDDR)
Under rare circumstances, the component may be susceptible to a failure mechanism that could impact device performance, including potential loss of function.
During the week of 15-Nov-2020, Medtronic initiated retrieval activities for 6 devices (5 US/1 OUS) for consignees in possession of potentially affected, non-implanted devices according to Medtronic records. Consignees were either verbally directed by Medtronic Field Representatives to quarantine product, or in the case of consigned units, the product was retrieved from consigned inventory and returned to Medtronic for engineering evaluation. The firm will follow with a letter to be disseminated on 12/04/2020 to the risk managers.
Distribution to US states of GA, PA, NJ, and France
2 units