FDA Recall Terminated

i-STAT DE handheld data processing module for clinical use, Software Version 2.8, List Number:08K46-01115200 (UDI: 00054749001255)

Recall: Z-0926-2018 · Initiated September 12, 2017

Recall

Recall Number
Z-0926-2018
Event Number
79150
Firm
Abbott Point Of Care Inc.
FEI Number
2245578
Product Code
JQP
Status
Terminated
Root Cause
Device Design
Initiated
September 12, 2017
Terminated
August 23, 2018
Address
400 College Rd E, Princeton, NJ, 08540-6607

Description

i-STAT DE handheld data processing module for clinical use, Software Version 2.8, List Number:08K46-01115200 (UDI: 00054749001255)

Reason

Issues resulting from upgrade to software version 2.8: (1) Location, operator, stored patient lists will not update, and (2) Customized Reference Ranges, Action Ranges, and Custom Reportable Ranges are reset to factory default values. No erroneous results are generated as a result of this issue.

Action

The following actions are recommended if you have already upgraded to version 2.8: 1. If you are using Location, Operator, or Patient lists. a. Delete the following files: i. C:\Istat32\LocInstData\LocInstData.txt ii. C:\Istat32\OpData\OpData.txt iii. C:\Istat32\PatData\PatData.txt iv. C:\Istat32\Log\iSTATDMI Errors.txt v. All files named C:\Istat32\Log\DeCustLogXX.txt; where XX is any two digits. b. Stop and restart the DECOMSRV service, which in turn restarts the DEDPSRV service. 2. If you are using customized Reference Ranges, Action Ranges, and Custom Reportable Ranges: a. Open DE Customization Workspace. b. For each profile, select preferences by clicking the button with the profile name. c. Navigate to the Results tab. d. Verify the Action and Custom Reportable Ranges. Click Edit to make changes. e. If changes are made, click the OK button to save. If you have not upgraded to i-STAT/DE version 2.8, make arrangements with your data management vendor to update to the newer version of i-STAT/DE at your earliest opportunity. DO NOT UPGRADE without coordinating with your data management vendor so that continuity of your i-STAT testing is not impacted. How should you verify what version of i-STAT/DE is installed on a computer? Using the Program and Features app in Windows is the best way to determine the version of i-STAT/DE installed on your system. To do so, follow these steps: 1. Open Control Panel and the Program and Features applet. 2. Locate the program named i-STAT/DE Version 2.8 in the list. 3. The version column will show 2.8 for the old version and 2.8.0.1 for the new version. If you determine the version 2.8.0.1 is not installed, make arrangements with your data management vendor to update to the latest validated version compatible with your data management system. Enclosed is an updated i-STAT/DE version 2.8, version qualifier, 2.8.0.1 CD-ROM and a Business Reply Card (BRC). Please complete and return the Business Reply c

Distribution

Distributed in 48 States: AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, the District of Columbia, Australia, Italy, Sweden, and Canada.

Quantity

894