Medtronic O-arm Imaging System Product Usage: image-intensified fluoroscopic x-ray system, mobile
Recall
- Recall Number
- Z-0926-2013
- Event Number
- 64256
- Firm
- Medtronic Navigation, Inc.
- FEI Number
- 3004785967
- Product Code
- OXO
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- January 31, 2013
- Posted
- March 8, 2013
- Terminated
- July 6, 2016
- Address
- 300 Foster St, Littleton, MA, 01460-2017
Description
Medtronic O-arm Imaging System Product Usage: image-intensified fluoroscopic x-ray system, mobile
Potential failure of the braking system that controls the O-arm Imaging System gantry movement in the vertical direction
Medtronic Navigation issued a Field Safety Notice letter dated January 18, 2013 to all affected customers. The letter identified the affected product, problem, and actions to be taken. The letter instructed customers not to discontinue use and that a Medtronic Navigation Representative will be contacting them shortly to schedule a time to conduct the motion control box upgrade. For questions contact your local Medtronic Navigation representative or call Technical Services Group at 800-595-9709 or (720) 890-3200, or email at rs.navtechsupport@ medtronic.com.
Worldwide Distribution - USA Nationwide and the countries of: Canada, AUSTRIA, BELGIUM, BRAZIL, BULGARIA, CHILE, CHINA, CROATIA, CZECH REPUBLIC, DENMARK,FINLAND, FRANCE, GERMANY, GREECE, INDIA, IRELAND, ISRAEL, ITALY, JAPAN ,KOREA, REPUBLIC OF KUWAIT, LEBANON, LIBYAN ARAB JAMAHIRIYA, MEXICO, NETHERLANDS, New Zealand, POLAND, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, Taiwan(R.O.C.) , THAILAND, TURKEY, UNITED ARAB EMIRATES. and UNITED KINGDOM.
463 units