FDA Recall
Terminated
StatLock Tandem Microbore Catheter Extension Set with preattached Clearlink Valve 6'' LL. Model number 2N9205K.
Recall: Z-0925-03
·
Initiated May 12, 2003
Recall
- Recall Number
- Z-0925-03
- Event Number
- 26365
- Firm
- Venetec Intl Inc
- FEI Number
- 1000117366
- Product Code
- FPA
- Status
- Terminated
- Root Cause
- Other
- Initiated
- May 12, 2003
- Posted
- June 12, 2003
- Terminated
- October 14, 2003
- Address
- 12555 High Bluff Dr, Ste 170, San Diego, CA, 92130
Description
StatLock Tandem Microbore Catheter Extension Set with preattached Clearlink Valve 6'' LL. Model number 2N9205K.
Reason
Set tubing disconnected from the male luer lock during use on patients. Misassembled.
Action
A recall letter was sent to the distributor who then sent thier own letters to the affected hospitals. Hospitals were reached on or about May 14, 2003 whereas the distributor was contacted by telephone and letter on or about May 12, 2003. It was verbally agreed that the distributor would subrecall the product.
Distribution
One distributor sent the product to 6 hospitals in the states of CT, IN, PA, TN.
Quantity
3150 units