Mizuho ProFx 6850 Operating Room Table, with orthopedic accessories, AC Powered, manufactured by Mizuho OSI, Union City, CA.
Recall
- Recall Number
- Z-0924-2011
- Event Number
- 56859
- Firm
- Mizuho Orthopedic Systems Inc
- FEI Number
- 2921578
- Product Code
- JEA
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- March 30, 2010
- Posted
- January 20, 2011
- Terminated
- May 17, 2011
- Address
- 30031 Ahern Ave, Union City, CA, 94587
Description
Mizuho ProFx 6850 Operating Room Table, with orthopedic accessories, AC Powered, manufactured by Mizuho OSI, Union City, CA.
Device component spar gear box locks up and may result in failure of push pull tube, which may result in release of breaking and impact.
All customers will be contacted to arrange for removal and shipping arrangements. Verification will be done by the Repair and Verification instructions for the leg spars, via registered letters. *******UPDATED INFO---The firm sent another notification letter---dated 12/22/10---which states in part: "****The upgrade involves the correction of your 6850 PROfx spars (6850-2100) at your facility by a Qualified Technician. The spars shall be upgraded to remove the 6850-2123, Drive Bracket, Spar Push Rod and replace it with the 6850-2170, Drive Bracket, Spar Push Rod. Upon completion of the upgrade, the 6850-2123, Drive Bracket, Spar Push rod shall be returned to Mizuho OSI for processing.*******
Worldwide Distribution -- USA & Germany.
17 units