IMS Patient Scale used on Sunrise Medical's Hoyer Presence and Stature Patient Lifts (Cradle), Models: HOY-4pt-WSC and HOY-StatureWSC
Recall
- Recall Number
- Z-0921-2007
- Event Number
- 37774
- Firm
- Integrated Measurement Systems, Inc.
- FEI Number
- 3002780774
- Product Code
- FRW
- Status
- Terminated
- Root Cause
- Other
- Initiated
- January 30, 2007
- Posted
- June 19, 2007
- Terminated
- March 2, 2008
- Address
- 440 Bonnie Ln, Elk Grove Village, IL, 60007-1908
Description
IMS Patient Scale used on Sunrise Medical's Hoyer Presence and Stature Patient Lifts (Cradle), Models: HOY-4pt-WSC and HOY-StatureWSC
Patient Scale may become detached, which can cause a patient to fall and result in injury.
The scale manufacturer, IMS, in conjunction with the lift manufacturer, Joerns Healthcare (previously Sunrise Medical), will initiate a field fix strategy consisting of shipping replacement 4-point cradles without scales, for use until a replacement can be designed, tested and validated. Joerns will utilize their Customer Service Help Line (1-800-826-0270) on behalf of IMS, to inform customers of the voluntary field correction, address any questions regarding the IMS scale field fix strategy and to provide additional information on return of affected IMS Patient Scales. Customers were notified by telephone beginning on January 30, 2007.
Worldwide
149 units