IMS Patient Scale used on Sunrise Medical's Hoyer Presence and Stature Patient Lifts (Cradle), Models: HOY-4pt-WSC and HOY-StatureWSC

Recall: Z-0921-2007 · Initiated January 30, 2007

Recall

Recall Number
Z-0921-2007
Event Number
37774
Firm
Integrated Measurement Systems, Inc.
FEI Number
3002780774
Product Code
FRW
Status
Terminated
Root Cause
Other
Initiated
January 30, 2007
Posted
June 19, 2007
Terminated
March 2, 2008
Address
440 Bonnie Ln, Elk Grove Village, IL, 60007-1908

Description

IMS Patient Scale used on Sunrise Medical's Hoyer Presence and Stature Patient Lifts (Cradle), Models: HOY-4pt-WSC and HOY-StatureWSC

Reason

Patient Scale may become detached, which can cause a patient to fall and result in injury.

Action

The scale manufacturer, IMS, in conjunction with the lift manufacturer, Joerns Healthcare (previously Sunrise Medical), will initiate a field fix strategy consisting of shipping replacement 4-point cradles without scales, for use until a replacement can be designed, tested and validated. Joerns will utilize their Customer Service Help Line (1-800-826-0270) on behalf of IMS, to inform customers of the voluntary field correction, address any questions regarding the IMS scale field fix strategy and to provide additional information on return of affected IMS Patient Scales. Customers were notified by telephone beginning on January 30, 2007.

Distribution

Worldwide

Quantity

149 units