Roche VALP2, Online TDM Valproic acid , for use with the Roche/Hitachi Cobas c 501 analyzer; Catalog No.04491041190.

Recall: Z-0920-2007 · Initiated April 30, 2007

Recall

Recall Number
Z-0920-2007
Event Number
37796
Firm
Roche Diagnostics Corp.
FEI Number
1823260
Product Code
LEG
Status
Terminated
Root Cause
Other
Initiated
April 30, 2007
Posted
June 9, 2007
Terminated
December 17, 2007
Address
9115 Hague Rd, Indianapolis, IN, 46250-0416

Description

Roche VALP2, Online TDM Valproic acid , for use with the Roche/Hitachi Cobas c 501 analyzer; Catalog No.04491041190.

Reason

Probe Wash: Reported results may have a positive bias because inadequate analyzer probe wash instructions may result in carry-over from the following assays: Creatinine Kinase L and Creatine Kinase-MBL.

Action

Consignees were issued a recall letter dated 4/30/07 instructing them to perform a R2 reagent probe wash when using CKL and CKMBL with the vancomycin or valproic acid assays.

Distribution

Nationwide.

Quantity

992