MUSE Cardiology Information System, Model Numbers: a) 3101011-040, b) 3101010-401, c) 3101010-009, d) 3101010-008
Recall
- Recall Number
- Z-0919-2022
- Event Number
- 89834
- Firm
- GE Healthcare, LLC
- FEI Number
- 2126677
- Product Code
- DQK
- Status
- Completed
- Root Cause
- Software Manufacturing/Software Deployment
- Initiated
- March 16, 2022
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
MUSE Cardiology Information System, Model Numbers: a) 3101011-040, b) 3101010-401, c) 3101010-009, d) 3101010-008
Two scenarios may cause edits to measurements and diagnosis statements can be lost after a test is Signed in the MUSE NX web client.
GE Healthcare notified its consignees by letter titled URGENT MEDICAL DEVICE CORRECTION on March 16, 2022. The letter explained the two scenarios which would produce the loss of edits and provided recommendations and a workaround. The firm will correct all affected products at no cost to you. The firm will be issuing a software patch. A GE Healthcare representative will contact affected customers to arrange for the correction. If you have any questions or concerns regarding this notification, contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative
US and Australia, Austria, Bahrain, Belgium, Brazil, Canada, Denmark, England, France, Germany, Ireland, Italy, Korea, Malaysia, Netherlands, Norway, Poland, Qatar, Saudi Arabia, Singapore, Spain, Sweden, Taiwan, Turkey
147 units