FDA Recall
Terminated
Carroll Healthcare brand Arro bed models BA3180-151 AE, BA3180-151 PE, BA3180-151 FE, BA3180-171 AE, BA3180-171 FE and BA3180-171 FE.
Recall: Z-0919-05
·
Initiated May 17, 2005
Recall
- Recall Number
- Z-0919-05
- Event Number
- 32093
- Firm
- Carroll Healthcare Inc. 1881 Huron St London Canada N6E1P5 Ontario
- FEI Number
- 3002706498
- Product Code
- FNL
- Status
- Terminated
- Root Cause
- Other
- Initiated
- May 17, 2005
- Posted
- June 22, 2005
- Terminated
- November 17, 2005
Description
Carroll Healthcare brand Arro bed models BA3180-151 AE, BA3180-151 PE, BA3180-151 FE, BA3180-171 AE, BA3180-171 FE and BA3180-171 FE.
Reason
The diodes in the control panel may fail, which could result in the thermal breakdown of the circuit board inside of the control box, and cause the bed to cease functioning.
Action
Consignees were sent a fax on 5/17/05 advising them of the problem and instructing them to lock-out the HILO pendant function on the attendant control pad until the firm visits their facility and corrects their beds.
Distribution
Nationwide.