FDA Recall Terminated

Carroll Healthcare brand Arro bed models BA3180-151 AE, BA3180-151 PE, BA3180-151 FE, BA3180-171 AE, BA3180-171 FE and BA3180-171 FE.

Recall: Z-0919-05 · Initiated May 17, 2005

Recall

Recall Number
Z-0919-05
Event Number
32093
Firm
Carroll Healthcare Inc. 1881 Huron St London Canada N6E1P5 Ontario
FEI Number
3002706498
Product Code
FNL
Status
Terminated
Root Cause
Other
Initiated
May 17, 2005
Posted
June 22, 2005
Terminated
November 17, 2005

Description

Carroll Healthcare brand Arro bed models BA3180-151 AE, BA3180-151 PE, BA3180-151 FE, BA3180-171 AE, BA3180-171 FE and BA3180-171 FE.

Reason

The diodes in the control panel may fail, which could result in the thermal breakdown of the circuit board inside of the control box, and cause the bed to cease functioning.

Action

Consignees were sent a fax on 5/17/05 advising them of the problem and instructing them to lock-out the HILO pendant function on the attendant control pad until the firm visits their facility and corrects their beds.

Distribution

Nationwide.