FDA Recall Open, Classified

Stryker MV3 bariatric bed, Part Number 5900000001

Recall: Z-0918-2026 · Initiated October 31, 2025

Recall

Recall Number
Z-0918-2026
Event Number
97953
Firm
Stryker Medical Division of Stryker Corporation
FEI Number
1831750
Product Code
OSI
Status
Open, Classified
Root Cause
Release of Material/Component prior to receiving test results
Initiated
October 31, 2025
Posted
December 11, 2025
Address
3800 E Centre Ave, Portage, MI, 49002-5826

Description

Stryker MV3 bariatric bed, Part Number 5900000001

Reason

Stryker has issued a voluntary recall of certain MV3 beds and Arise 1000EX mattresses because some of these products have been sold in close proximity, suggesting a potential intent for them to be used together. These products have not been tested for compatibility and thus are not designed to be used together. The combination of both is not recommended by Stryker.

Action

Stryker issued an URGENT: VOLUNTARY MEDICAL DEVICE SAFETY notice to its sales representatives on 11/13/2025 for hand delivery to its consignees on 11/14/2025. The notice explained the issue, potential hazard, and requested the following: "Actions needed: 1. Locate the MV3 and Arise 1000EX devices that have been shipped to your facility. Serial numbers are provided on the enclosed business reply form. 2. If an MV3 is being used together with an Arise 1000EX mattress, Stryker recommends the use to be discontinued upon receipt of this notification. 3. Complete the enclosed business reply form to confirm receipt of this notification. 4. Return your completed business reply form to Stryker by fax +1 269 488 8691 or email [email protected]. 5. For those requesting a replacement of a compatible mattress, Stryker will contact you upon receipt of your Business Reply Form for next steps. 6. If you have loaned or sold any of the products listed in this letter, please forward a copy of this notice to the new users and advise us of their new location." For questions or concerns, contact Customer Service at +1 800 327 0770, Option 1 for Customer Support, Monday through Friday from 8:00 a.m. to 7:00 p.m. ET.

Distribution

US, nationwide

Quantity

294 units