FDA Recall Open, Classified

Portrait Core Services, a set of software services that enable the communication and interaction of the Portrait Mobile Monitoring Solution components and will integrate into existing healthcare facility infrastructure and clinical information systems, Model CSSXB

Recall: Z-0918-2025 · Initiated December 6, 2024

Recall

Recall Number
Z-0918-2025
Event Number
96027
Firm
GE Medical Systems, LLC
FEI Number
2126677
Product Code
MWI
Status
Open, Classified
Root Cause
Software design
Initiated
December 6, 2024
Posted
January 15, 2025
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

Portrait Core Services, a set of software services that enable the communication and interaction of the Portrait Mobile Monitoring Solution components and will integrate into existing healthcare facility infrastructure and clinical information systems, Model CSSXB

Reason

There is a software issue that can result in the loss of patient monitoring on the Portrait Central Viewer Application after 425 days of continuous run-time.

Action

An URGENT MEDICAL DEVICE CORRECTION notification letter dated 12/6/24 was sent to customers. Actions to be taken by Customer/User: You can continue to use your Portrait" Mobile Monitoring Solution v1.1 product, since all potentially impacted systems were recently installed and the safety issue does not manifest until after 425 days of continuous use. Please follow the annual Maintenance Schedule described in Service Manual (PN 5929144) for your Portrait" Mobile to reboot the system. Rebooting the system by following this procedure resets the continuous days of use to zero. Please complete and return the attached Acknowledgement Form to [email protected]. After the Portrait Core Services v1.1 software is updated and corrected by GE HealthCare, please discontinue usage of any Portrait Core Services v1.1 software versions 1.1.0 or 1.1.1 and destroy any software media containing those versions. GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

Distribution

Worldwide - US Nationwide distribution in the states of OH, TX and the countries of Australia, China, Czech Republic, France, Italy, Netherlands, Qatar, Singapore, UAE, UK.

Quantity

7 systems