FDA Recall Open, Classified

Cardiosave Rescue. Intra-Aortic Balloon Pump system.

Recall: Z-0917-2026 · Initiated November 4, 2025

Recall

Recall Number
Z-0917-2026
Event Number
97946
Firm
Datascope Corp.
FEI Number
3001418283
Product Code
DSP
Status
Open, Classified
Root Cause
Component design/selection
Initiated
November 4, 2025
Posted
December 11, 2025
Address
1300 Macarthur Blvd, Mahwah, NJ, 07430-2052

Description

Cardiosave Rescue. Intra-Aortic Balloon Pump system.

Reason

The IFU addendum revises the Preventative Maintenance schedule to align with the update introduced in the Service Manual as of June 2023.

Action

An URGENT MEDICAL DEVICE CORRECTION NOTICE, dated October 2025, was sent via FEDEX to consignees. The notification notifies consignees of four addendum updates: 1) NVRAM Preventive Maintenance (New FSCA - reflected in this recall entry), 2) Vibration and Shock Table Specifications (New FSCA), 3) Helium O-Ring Preventive Maintenance (FSCA 2249723-01/24/2023-004-C), and 4) Battery Runtime Specifications (FSCA 2249723-09/10/2021-001-R). The notification instructs consignees regarding the NVRAM Preventive Maintenance to review the provided IFU addendum, update local preventive maintenance procedures as such, and ensure all NVRAM batteries are within their life expectancy. If a NVRAM is greater than 8 years old, until it can be replaced, it is advised a back-up Cardiosave IABP be available to provide therapy. Consignees with any questions can contact Datascope/Getinge Customer Support at 1-88-943-8872 (options 4, 2, 1) Monday through Friday, from 8:00 AM EST and 6:00 PM EST.

Distribution

Domestic: Nationwide Distribution; Foreign: Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bosnia And Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, C¿te d Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Ghana, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Mauritius, Mexico, Montenegro, Morocco, Myanmar, Netherlands, New Zealand, Nicaragua, Nigeria, North Macedonia, Norway, Oman, Pakistan, Palestine Territory, Peru, Philippines, Poland, Portugal, Qatar, Russian Federation, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, Yemen, Zambia.

Quantity

47 units