Lotus TAVR 27mm, Transcatheter Aortic Valve Prosthesis Premounted on Delivery System; Sterile Material number H749LTV270, Catalog Number LTV27; Product Usage: The Lotus Valve System is intended to improve aortic valve function for symptomatic subjects with severe calcific aortic stenosis (aortic valve area [AVA] of <1.0 cm2 or AVA index of <0.6 em /m2) who are at extreme or high risk for standard surgical valve replacement.
Recall
- Recall Number
- Z-0917-2015
- Event Number
- 69931
- Firm
- Boston Scientific Corporation
- FEI Number
- 3003882783
- Product Code
- NPT
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- November 19, 2014
- Posted
- January 13, 2015
- Terminated
- March 26, 2015
- Address
- 160 Knowles Dr, Los Gatos, CA, 95032-1828
Description
Lotus TAVR 27mm, Transcatheter Aortic Valve Prosthesis Premounted on Delivery System; Sterile Material number H749LTV270, Catalog Number LTV27; Product Usage: The Lotus Valve System is intended to improve aortic valve function for symptomatic subjects with severe calcific aortic stenosis (aortic valve area [AVA] of <1.0 cm2 or AVA index of <0.6 em /m2) who are at extreme or high risk for standard surgical valve replacement.
Lotus valve became unlocked during release from the delivery system. This may lead to percutaneous insertion of another valve, resulting in a Valve in Valve (ViV) implantation. If this occurs, the first valve could embolize and it may be necessary to convert the patient to surgery.
Boston Scientific notified consignees on November 19, 2014, of the problem via letter through overnight delivery directing consignees to segregate and return affected product to Boston Scientific utilizing the Field Safety Corrective Action (FSCA) Instructions provided. The FSCA details that consignees should: Immediately discontinue use of and segregate product; Complete and return the Account Reply Verification Tracking form and Package/Ship the affected product.
Distributed only in the countries of Finland, France, Germany, Great Britain, Italy, Norway, Spain, Sweden, and Switzerland.
278 units - total all sizes