Roche Online TDM Tobramycin, for use on Roche/Hitachi analyzers 911,912, 917 and Modular P; Catalog No. 04642511190. (Note: Not distributed within the United States)

Recall: Z-0917-2007 · Initiated April 5, 2007

Recall

Recall Number
Z-0917-2007
Event Number
37795
Firm
Roche Diagnostics Corp.
FEI Number
1823260
Product Code
KLB
Status
Terminated
Root Cause
Other
Initiated
April 5, 2007
Posted
June 9, 2007
Terminated
April 30, 2008
Address
9115 Hague Rd, Indianapolis, IN, 46256-1025

Description

Roche Online TDM Tobramycin, for use on Roche/Hitachi analyzers 911,912, 917 and Modular P; Catalog No. 04642511190. (Note: Not distributed within the United States)

Reason

Reported results may have a positive bias of up to 113% because inadequate cleaning instructions may result in carry-over of the following reagents: Online TDM carbamazepine, Online TDM gentamycin, Online TDM phenytoin, Online DAT cocaine II, Online DAT methadone II, Online DAT opiates II and Online DAT THC II.

Action

Foreign affiliates were notified via email letter dated 4/5/07 requesting that to notify their customers of the need to implement a special wash cycle to prevent reagent carry over.

Distribution

International Distribution ONLY-Australia, Canada and Germany.

Quantity

474 total (Z-0917-0918-2007)