FDA Recall Terminated

Cardiac Science Powerheart Automated External Defibrillator G3 9300A

Recall: Z-0916-2009 · Initiated October 13, 2008

Recall

Recall Number
Z-0916-2009
Event Number
50064
Firm
Cardiac Science Corporation
FEI Number
2112020
Product Code
MKJ
Status
Terminated
Root Cause
Software design
Initiated
October 13, 2008
Posted
February 25, 2009
Terminated
December 8, 2011
Address
3303 Monte Villa Pkwy, Bothell, WA, 98021-8969

Description

Cardiac Science Powerheart Automated External Defibrillator G3 9300A

Reason

Potential for AED to not deliver defibrillation therapy.

Action

On October 13, 2008 Cardiac Science began sending via Fed-Ex letters and software upgrade kits. Mailing of letters and software upgrade kits is anticipated to be completed by November 21, 2008. The letter instructs consignees to install the software update, assure that the AED information test file generated by the update software is returned to the firm, and complete the postage paid response card and return it to the firm. If the affected devices are no longer in the possession of the consignee, the consignee is to contact the firm and forward the update kit and letter to the person who has the AED in their possession. A separate letter is being sent to distributors. The letter advised to install the software upgrades for any units in stock or to forward the software upgrade along with the end user letter enclosed and that the AED information text file generated by the update software must be returned to the firm. The firm will visit the locations of individual national account customers and install the update. OEM partners will be notified and provided the information and update kits to provide to their customers.

Distribution

Worldwide Distribution: USA, Puerto Rico, Virgin Islands, Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czechoslovakia, Denmark, Dominican Republic, France, Germany, Greece, Guam, Hong Kong, Hungary, Ireland, Israel, Italy, Jamaica, Japan, Kenya, Kuwait, Libya, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, UK, and Ukraine.

Quantity

9,633 devices