AMSCO C and AMSCO 400 Steam Sterilizers designed for sterilization of heat and moisture-stable materials used in healthcare facilities.
Recall
- Recall Number
- Z-0915-2013
- Event Number
- 64440
- Firm
- Steris Corporation
- FEI Number
- 1527821
- Product Code
- FLE
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- February 12, 2013
- Posted
- March 6, 2013
- Terminated
- December 5, 2013
- Address
- 5960 Heisley Rd, Mentor, OH, 44060-1834
Description
AMSCO C and AMSCO 400 Steam Sterilizers designed for sterilization of heat and moisture-stable materials used in healthcare facilities.
The control boards alarm set points and temperature tolerances allowed for erroneous activation of the AD (analog/digital) alarm as temperature fluctuated during operation. When this alarm occurs, the sterilizer locks up and cannot be used until serviced, causing Customer nuisance.
Steris Corporation initiated their recall on February 12, 2013. They will notify customers via onsite visits by STERIS Field Service Technicians to each affected Customer site. For further questions, please call (440) 392-7601.
Worldwide Distribution - USA including Canada and the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, SK, TN, TX, UT, VA, VT, WA, WI, and WV.
395 Units