FDA Recall Terminated

AMSCO C and AMSCO 400 Steam Sterilizers designed for sterilization of heat and moisture-stable materials used in healthcare facilities.

Recall: Z-0915-2013 · Initiated February 12, 2013

Recall

Recall Number
Z-0915-2013
Event Number
64440
Firm
Steris Corporation
FEI Number
1527821
Product Code
FLE
Status
Terminated
Root Cause
Software design
Initiated
February 12, 2013
Posted
March 6, 2013
Terminated
December 5, 2013
Address
5960 Heisley Rd, Mentor, OH, 44060-1834

Description

AMSCO C and AMSCO 400 Steam Sterilizers designed for sterilization of heat and moisture-stable materials used in healthcare facilities.

Reason

The control boards alarm set points and temperature tolerances allowed for erroneous activation of the AD (analog/digital) alarm as temperature fluctuated during operation. When this alarm occurs, the sterilizer locks up and cannot be used until serviced, causing Customer nuisance.

Action

Steris Corporation initiated their recall on February 12, 2013. They will notify customers via onsite visits by STERIS Field Service Technicians to each affected Customer site. For further questions, please call (440) 392-7601.

Distribution

Worldwide Distribution - USA including Canada and the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, SK, TN, TX, UT, VA, VT, WA, WI, and WV.

Quantity

395 Units