FDA Recall Terminated

Access Peristaltic Pump component of: UniCel DxI 800 Access Immunoassay Systems, UniCel DxI 600 Access Immunoassay Systems, UniCel DxC 880i SYNCHRON Access Clinical Systems. Intended to be used with the Access Immunoassay System to remove waste from the system.

Recall: Z-0914-2010 · Initiated October 9, 2008

Recall

Recall Number
Z-0914-2010
Event Number
50703
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
JJE
Status
Terminated
Root Cause
Other
Initiated
October 9, 2008
Posted
March 2, 2010
Terminated
May 4, 2012
Address
200 S Kraemer Blvd, Brea, CA, 92821-6208

Description

Access Peristaltic Pump component of: UniCel DxI 800 Access Immunoassay Systems, UniCel DxI 600 Access Immunoassay Systems, UniCel DxC 880i SYNCHRON Access Clinical Systems. Intended to be used with the Access Immunoassay System to remove waste from the system.

Reason

The recall was initiated after Beckman Coulter confirmed reports of a new premature failure of the upper aspirate peristaltic pump in the systems listed above. Premature failure of the pump may negatively affect precision. The failure is limited to those UniCel DXI 800, DXI 600, and DxC 880i systems which have a pump identified with one of the new pump manufacturing numbers (Mfg. Nos.) listed ab

Action

Beckman Coulter contacted consignees via "Urgent Product Corrective Action" letter dated October 16, 2008 to inform them of the recall and to arrange a date to service the affected product as soon as possible. For further information, contact Beckman Coulter: (a)In the U.S.A. or Canada, contact Technical Support at 800-854-3633 or online at www.beckman.com/customersupport. (b) Outside the U.S.A. and Canada, contact your technical support representative.

Distribution

Worldwide Distribution -- United States and Canada.

Quantity

107 (US), 7 (OUS)