EP-4 Computerized Stimulator -- 2 channel and 4 channel. The product consists of a touchscreen portable PC controlling a two channel dedicated EP stimulator. The stimulator is capable of single channel, simultaneous and sequential stimulation at programmed output settings using built-in protocols, and includes an emergency stimulation function.
Recall
- Recall Number
- Z-0914-06
- Event Number
- 35206
- Firm
- EP Medsystems
- FEI Number
- 2126673
- Product Code
- JOQ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 4, 2006
- Posted
- May 31, 2006
- Terminated
- April 24, 2007
- Address
- 575 Route 73 North, Bldg D, Cooper Run Executive Park West Berlin, NJ, 08091
Description
EP-4 Computerized Stimulator -- 2 channel and 4 channel. The product consists of a touchscreen portable PC controlling a two channel dedicated EP stimulator. The stimulator is capable of single channel, simultaneous and sequential stimulation at programmed output settings using built-in protocols, and includes an emergency stimulation function.
Under certain circumstances the device may exhibit the following behaviors: 1)-Initiation of a rapid stimulation train without command, when the device is in use in conjuction with an electrosurgical unit or RF ablation unit-2)-Presence of a DC offset voltage on an output channel-3)-Inability to synchronize with a non-EP MedSystems recording system
The firm initiated the recall on April 4, 2006 through a notification letter sent to the electrophysiology lab managers at the healthcare institutions and two domestic distributors to which the device has been distributed. The firm used a traceable courier service to deliver notification of the recall.
Product was shipped to hospitals/users throughout the US. The firm has a Veterans Affair contract number V797P-4650A. Also the firm has shipped the devices to Germany, United Kingdom, Netherlands, Italy, Japan, France, Saudi Arabia, Hong Kong, Syria, Malaysia, Sweden, Armenia, Australia, Canada, Iceland, S. Africa, and Taiwan.
187 units