Giraffe Infant Warmers, part number M1118179, labeled in part *** GE Healthcare 8880 Gorman Road Laurel, MD 20723***www.gehealthcare.com* Giraffe Infant Warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology.
Recall
- Recall Number
- Z-0912-2014
- Event Number
- 67225
- Firm
- GE Healthcare
- FEI Number
- 1121732
- Product Code
- FMT
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- December 12, 2013
- Posted
- February 3, 2014
- Terminated
- June 1, 2015
- Address
- 8880 Gorman Rd, Laurel, MD, 20723-5800
Description
Giraffe Infant Warmers, part number M1118179, labeled in part *** GE Healthcare 8880 Gorman Road Laurel, MD 20723***www.gehealthcare.com* Giraffe Infant Warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology.
Medical device adjustment knob may be loose and affect the accuracy of blended Air/Oxygen gas mixture to patients.
GE Healthcare notified end users with a letter flagged as Urgent Medical Device Correction letter on 12/12/2013. The notification advised that GE Healthcare has become aware of a potential safety issue associated with the loosening of the blender knob shaft on all Resuscitation units. The firm advised that accounts ensure that all potential users in their facility are made aware of this safety notification and the recommended actions described in the notification: 1. Perform blender accuracy check as outlined in Section 2.5-2.5.1 of the Service Manual to verify that the O2 concentration set at 21% is within 16-26% O2 and the concentration at 100% setting is within the recommended range (95-105% O2). 2. If the blender accuracy test results do not meet specifications, remove the unit from patient use and contact a GE Healthcare Service Representative at the number listed below. 3. If the blender accuracy check passes in Step 1 above, you may continue to use your resuscitation system. The following precautions are recommended to ensure blender accuracy is maintained between system checks. Use an independent oxygen analyzer, as determined by your hospital policy, to check blended Air/O2 accuracy. When delivering Oxygen, use a pulse oximeter to monitor your patient. Customers were requested to contact a GE Healthcare Service Representative at 1-800-345-2700 to arrange for correction if the device is found to be defective.
Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Uruguay, Venezuela, Albania, Algeria, Australia, Austria, Bahrain, Belarus, Belgium, Bolivia, Canada, Chile, China, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Kazakhstan, Korea, Kuwait, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Arab Emirates, and United Kingdom.
1970 units