FDA Recall Terminated

Atricure Inc., Max1 Isolator Transpolar Pen. Model A000163. Used to ablate cardiac tissues during cardiac surgery using radiofrequency energy when connected to the AtriCure Ablation and Sensing Unit or for temporary cardiac pacing, sensing, recording, and stimulation during the evaluation of cardiac arrhythmias during surgery when connected to a temporary external cardiac pacemaker or recording device.

Recall: Z-0912-2013 · Initiated January 14, 2013

Recall

Recall Number
Z-0912-2013
Event Number
64404
Firm
Atricure Inc
FEI Number
3011706110
Product Code
OCL
Status
Terminated
Root Cause
Employee error
Initiated
January 14, 2013
Posted
March 6, 2013
Terminated
June 11, 2013
Address
6217 Centre Park Drive West, Chester, OH, 45069

Description

Atricure Inc., Max1 Isolator Transpolar Pen. Model A000163. Used to ablate cardiac tissues during cardiac surgery using radiofrequency energy when connected to the AtriCure Ablation and Sensing Unit or for temporary cardiac pacing, sensing, recording, and stimulation during the evaluation of cardiac arrhythmias during surgery when connected to a temporary external cardiac pacemaker or recording device.

Reason

Products from lot no. 41865 may have been sent out with Incorrect Instructions for Use (IFU) labeling.

Action

AtriCure sent a Urgent Medical Device Recall/Correction letter dated January 14, 2013, to all affected customers. The letter identified the product the problem and the action to be taken by the customer. AtriCure has identified that products from lot no. 41865 may have been sent out with Incorrect Instructions for Use (IFU) labeling. If the MAX1 is used as intended there is no risk to the patient as a result of this issue. The correct instructions for Use for the MAX1 Pen is included with this notification. AtriCure will replace any affected product if requested by the facility. Please review and fill out the attached " Acknowledgement Form" and either fax to AtriCure at (513) 644-4753. If you need to contact customer service, their number is 1-866-349-2342.

Distribution

Nationwide Distribution including the states of MD, PA and TX.

Quantity

5 devices