FDA Recall Terminated

Varian Medical Systems, Eclipse Client software versions with build numbers from 7.3.10 through 8.6.15 (i.e. excluding 8.6.17); Model Number: H48; Radiation Treatment Planning System

Recall: Z-0911-2011 · Initiated August 17, 0010

Recall

Recall Number
Z-0911-2011
Event Number
56772
Firm
Varian Medical Systems, Palo Alto, CA 911 Hansen Way Palo Alto CA 94304-1028
FEI Number
2916710
Product Code
MUJ
Status
Terminated
Root Cause
Software design
Initiated
August 17, 0010
Posted
January 18, 2011
Terminated
July 21, 2011

Description

Varian Medical Systems, Eclipse Client software versions with build numbers from 7.3.10 through 8.6.15 (i.e. excluding 8.6.17); Model Number: H48; Radiation Treatment Planning System

Reason

Possible incorrect dose calculation: An anomaly in certain Eclipse versions where, under certain specific sequences of user actions, the dose distribution may fail to be re-calculated even though changes have been made which affect dose.

Action

Varian Medical Systems Urgent Medical Device Correction letters were sent to all affected users on August 6, 2010. Letters were also sent to Varian Sales, Marketing and Service organizations to inform them of the recall.

Distribution

Worldwide Distribution.

Quantity

7019