FDA Recall Terminated

Medtronic Paceart 2004 Second Edition & Paceart 2005 First Edition versions of the Paceart Gateway Software. The Medtronic Paceart System is a personal computer based pacemaker-testing system, a 12-lead electrocardiograph, and a transtelephonic receiving station. It provides data that can be used to analyze implanted pacemaker performance based on electrocardiographic measurements, either taken directly from the patient or programmer. The system can measure, store, and display any of the 12 standard leads. Reports and charts are available by means of a laser printer. The system also includes a database that collects and stores patient data. Optional software allows the device to function as an unattended transtelephonic receiving station.

Recall: Z-0911-06 · Initiated February 2, 2006

Recall

Recall Number
Z-0911-06
Event Number
35141
Firm
Medtronic Inc. Cardiac Rhythm Managment
FEI Number
2182208
Product Code
DPS
Status
Terminated
Root Cause
Other
Initiated
February 2, 2006
Posted
May 31, 2006
Terminated
December 31, 2006
Address
7000 Central Ave Ne, Minneapolis, MN, 55432-3568

Description

Medtronic Paceart 2004 Second Edition & Paceart 2005 First Edition versions of the Paceart Gateway Software. The Medtronic Paceart System is a personal computer based pacemaker-testing system, a 12-lead electrocardiograph, and a transtelephonic receiving station. It provides data that can be used to analyze implanted pacemaker performance based on electrocardiographic measurements, either taken directly from the patient or programmer. The system can measure, store, and display any of the 12 standard leads. Reports and charts are available by means of a laser printer. The system also includes a database that collects and stores patient data. Optional software allows the device to function as an unattended transtelephonic receiving station.

Reason

The Paceart System could inappropriately insert data from a patient's Medtronic CareLink transmission into another patient's Paceart record. An issue associated with the automated batch transfer of implanted cardiac device data from the Medtronic CareLink network to the Paceart System exists.

Action

Affected user and potential users of the Gateway product were notified via phone and/ or a customer notification. The notification offers interim options until a software update is available. Customers are asked to respond back to Metronic with an enclosed Acknowledgement form.

Distribution

Nationwide

Quantity

210