Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required) base, Terumo Cardiovascular Systems, Ann Arbor, MI. Model 801764. (Not distributed within the United States). The device is indicated for use in the extracorporeal circulation of blood for arterial perfusion and cardiopulmonary bypass procedures by a perfusionist.
Recall
- Recall Number
- Z-0910-2009
- Event Number
- 50554
- Firm
- Terumo Cardiovascular Systems Corporation
- FEI Number
- 1828100
- Product Code
- DTQ
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- November 7, 2008
- Posted
- February 24, 2009
- Terminated
- July 7, 2009
- Address
- 6200 Jackson Road, Ann Arbor, MI, 48103-9586
Description
Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required) base, Terumo Cardiovascular Systems, Ann Arbor, MI. Model 801764. (Not distributed within the United States). The device is indicated for use in the extracorporeal circulation of blood for arterial perfusion and cardiopulmonary bypass procedures by a perfusionist.
The power manager board may cause a short circuit, resulting in loss of system power, loss of central control monitor or electrical arcing.
An "Urgent Medical Device Correction" letter was sent to consignees on November 7, 2008. The letter described the reason for correction and provided instructions to customers. Terumo Cardiovascular Systems Corporation will replace the power manager boards on the affected devices. Customers were instructed to complete and fax the Customer Response Form to CVS Customer Service at 1-800-292-6551 (US only) or respond by telephone at 1-800-521-2818. Please direct questions to CVS Customers Service at 1-800-521-2818.
Worldwide Distribution including USA and countries of Australia, Belgium, Canada, Columbia, Costa Rica, Dominican Republic, Egypt, Germany, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, United Arab Emirates and Viet Nam.
515