FDA Recall Terminated

Philips Easy Diagnost Eleva

Recall: Z-0910-2007 · Initiated July 23, 2007

Recall

Recall Number
Z-0910-2007
Event Number
44910
Firm
Philips Medical Systems North America Co. Phillips
FEI Number
1000524572
Product Code
KPR
Status
Terminated
Root Cause
Other
Initiated
July 23, 2007
Posted
September 25, 2007
Terminated
September 17, 2012
Address
22100 Bothell Everett Hwy, Bothell, WA, 98021-8431

Description

Philips Easy Diagnost Eleva

Reason

Units do not comply with the requirement to provide automatic beam limitation within +/- 3 degrees of horizontal, vertical, and perpendicular to the plane of the image receptor during radiographic procedures.

Action

On 7/23/07 Philips mailed a letter dated July 9, 2007 titled Electronic Product Radiation Warning. The letter advises customers of the issue and asks that customers apply the enclosed cautionary label to the device. The letter also explains that there will be a system modification to be installed. On 8/20/07 the firm sent the same letter, along with a letter dated 8/20/07 to customers thought to have either scrapped their units and the customer had previously indicated they were no longer using the unit. The 8/20/07 letter requested the customer contact Philps if they are currently using the device.

Distribution

Units were distributed to 1543 hospitals and medical centers throughout the US. No international distribution from this firm.

Quantity

289 units