CONTAK RENEWAL 2 (model H155) Heart Failure Device. Sterilized with gaseous ethylene oxide. Guidant corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. Renewal provides ventricular tachyarrhythmia and cardiac resynchronization therapies.

Recall: Z-0908-05 · Initiated June 17, 2005

Recall

Recall Number
Z-0908-05
Event Number
32380
Firm
Guidant Corporation
FEI Number
2124215
Product Code
NIK
Status
Terminated
Root Cause
Other
Initiated
June 17, 2005
Posted
June 30, 2005
Terminated
January 10, 2007
Address
4100 Hamline Ave N, Saint Paul, MN, 55112-5700

Description

CONTAK RENEWAL 2 (model H155) Heart Failure Device. Sterilized with gaseous ethylene oxide. Guidant corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. Renewal provides ventricular tachyarrhythmia and cardiac resynchronization therapies.

Reason

Laboratory analysis revealed that a deterioration in a wire insulator within the lead connector block, in conjunction with other factors, could cause a short circuit and loss of device function due to diversion of therapy energy away from the heart and into device circuitry.

Action

Dear Doctor letter was dated and issued 06/17/05 informing Doctors of an shorting issue of the Renewal 1 & 2 CRT-D devices. The letter gave indications of device failure and recommendations for patient visits. A press release was also issued on 06/17/05.

Distribution

worldwide. Including the United States and countries such as: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Cyprus, Czech Republic, Czechoslovakia, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Monaco, Netherlands, New Zealand, Norway, Portugal, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Thailand, Turkey, United Kingdom and Venezuela.

Quantity

16,000 (Approximately 11,900 remain implanted worldwide) - both products combined