CONTAK RENEWAL (model H135) Heart Failure Device Cardiac Resynchronization Therapy Defibrillators (CRT-Ds). Sterilized with gaseous ethylene oxide. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798. USA. Renewal provides ventricular tachyarrhythmia and cardiac resynchronization therapies.
Recall
- Recall Number
- Z-0907-05
- Event Number
- 32380
- Firm
- Guidant Corporation
- FEI Number
- 2124215
- Product Code
- NIK
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 17, 2005
- Posted
- June 30, 2005
- Terminated
- January 10, 2007
- Address
- 4100 Hamline Ave N, Saint Paul, MN, 55112-5700
Description
CONTAK RENEWAL (model H135) Heart Failure Device Cardiac Resynchronization Therapy Defibrillators (CRT-Ds). Sterilized with gaseous ethylene oxide. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798. USA. Renewal provides ventricular tachyarrhythmia and cardiac resynchronization therapies.
Laboratory analysis revealed that a deterioration in a wire insulator within the lead connector block, in conjunction with other factors, could cause a short circuit and loss of device function due to diversion of therapy energy away from the heart and into device circuitry.
Dear Doctor letter was dated and issued 06/17/05 informing Doctors of an shorting issue of the Renewal 1 & 2 CRT-D devices. The letter gave indications of device failure and recommendations for patient visits. A press release was also issued on 06/17/05.
worldwide. Including the United States and countries such as: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Cyprus, Czech Republic, Czechoslovakia, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Monaco, Netherlands, New Zealand, Norway, Portugal, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Thailand, Turkey, United Kingdom and Venezuela.
16,000 (Approximately 11,900 remain implanted worldwide)