FDA Recall
Terminated
Olympus, Model No. TJF-160F, flexible scope, large diameter, video duodenoscope
Recall: Z-0905-2016
·
Initiated January 20, 2016
Recall
- Recall Number
- Z-0905-2016
- Event Number
- 73088
- Firm
- Integrated Medical Systems Inc
- FEI Number
- 3010663403
- Product Code
- GCQ
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- January 20, 2016
- Terminated
- October 13, 2016
- Address
- 3200 2nd Ave N, Birmingham, AL, 35222-1212
Description
Olympus, Model No. TJF-160F, flexible scope, large diameter, video duodenoscope
Reason
If excessive pressure is applied to the distal end of the endoscope, the internal coil pipe assembly at the distal end of the endoscope can become detached and angulation may not operate as intended.
Action
The firm notified their consignees of the issue by Fed Ex on 01/20/2016. The firm plans to coordinate with the medical facilities for inspection of the endoscopes.
Distribution
Distributed US (nationwide) and in Australia.
Quantity
16 instruments