FDA Recall Terminated

BuckyDiagnost general purpose x-ray system with Optimus Generator. General Purpose Diagnostic Radiography.

Recall: Z-0905-2011 · Initiated January 17, 2011

Recall

Recall Number
Z-0905-2011
Event Number
58077
Firm
Philips Healthcare Inc.
FEI Number
1218950
Product Code
IZO
Status
Terminated
Root Cause
Radiation Control for Health and Safety Act
Initiated
January 17, 2011
Posted
March 4, 2011
Terminated
August 11, 2020
Address
3000 Minuteman Road, Andover, MA, 01810

Description

BuckyDiagnost general purpose x-ray system with Optimus Generator. General Purpose Diagnostic Radiography.

Reason

Affected systems were determined not to prevent run-on of a selected technique factor if the technique selector and the exposure control were activated simultaneously.

Action

A Customer Information BuckyDiagnost Diagnostic X-Ray System letter, dated January 6, 2011, was sent to customers. The letter identifies the affected product, the reason for the recall, actions that are to be taken by customers, and actions that are planned by the firm. Customers are advised to make sure that the desired exposure values are displayed and that they have released the technique button BEFORE they start the exposure by pressing the exposure switch whenever manually changing the exposure parameters and in order to avoid the problem. Philips will issue a Field Change Order (FCO) to resolve the issue by updating the system software free of charge. Philips will contact customers to implement the FCO. Further information or support concerning this issue can be obtained by contacting your local Philips representative. The Field Service Engineers began software implementation the week of March 21, 2011.

Distribution

Worldwide Distribution

Quantity

1810 units in the US