Virotek SAFEL*E*T 1.0 mm Safety Lancets, 1.0 mm depth, stock code 5000052, packaged in single unit blister packs, 150 or 200 per box, 6 boxes per carton; Virotek, L.L.C., 900 Asbury Dr., Buffalo Grove, IL 60089
Recall
- Recall Number
- Z-0904-05
- Event Number
- 31966
- Firm
- Efoora, Inc. dba Virotek, L.L.C.
- FEI Number
- 3002689104
- Product Code
- FMK
- Status
- Terminated
- Root Cause
- Other
- Initiated
- May 5, 2005
- Posted
- June 17, 2005
- Terminated
- September 2, 2008
- Address
- 900 Asbury Dr, Buffalo Grove, IL, 60089-4557
Description
Virotek SAFEL*E*T 1.0 mm Safety Lancets, 1.0 mm depth, stock code 5000052, packaged in single unit blister packs, 150 or 200 per box, 6 boxes per carton; Virotek, L.L.C., 900 Asbury Dr., Buffalo Grove, IL 60089
The sterility of the lancets may be compromised due to inadequate packaging seals.
Efoora/Virotek sent recall notifications to their customers via fax and DHL delivery on 5/5/05 through 5/27/05. The accounts were informed that the sterility of the lancets had been compromised due to inadequate seals. The accounts were requested to check their inventories for the affected lots, discontinue their use/distribution, and return the product to Efoora for a refund. Any questions were directed to Mary Hill, Director of Quality Assurance at 847-634-4500, ext. 277.
Nationwide and internationally to Australia, Belgium, Bolivia, Colombia, Finland, Israel, Italy, Japan, Mexico, Nigeria, Russia, Singapore, South Africa, Spain, Sweden, Taiwan, Trinidad, United Kingdom and Zimbabwe.
250,150 units