FDA Recall
Terminated
Hill-Rom brand Century+ series bed
Recall: Z-0903-03
·
Initiated April 24, 2003
Recall
- Recall Number
- Z-0903-03
- Event Number
- 25899
- Firm
- Hill-Rom, Inc.
- FEI Number
- 1824206
- Product Code
- KNC
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 24, 2003
- Posted
- May 29, 2003
- Terminated
- September 25, 2003
- Address
- 125 E Pearl St, Batesville, IN, 47006
Description
Hill-Rom brand Century+ series bed
Reason
Possible electrical shock hazard, as the power cord grounding pin may break off or become detached.
Action
Letters dated April 24, 2003 were sent to each consignee instructing them to inspect the beds for power cords with non-flared ground pins, to cut plug off any such cords when found, and replace these plugs with new plugs with the flared ground pin. [The type of plug on the unit can be seen by the naked eye, as the plug is transparent plastic molding.]
Distribution
United States, Bahamas, Bermuda, Canada, Colombia, Costa Rica, Dominican Republic, Guam, Japan, Mexico, Panama, Saudi Arabia, South Korea, Taiwan and Venezuela