Hill-Rom brand Affinity three birthing bed

Recall: Z-0902-03 · Initiated April 24, 2003

Recall

Recall Number
Z-0902-03
Event Number
25899
Firm
Hill-Rom, Inc.
FEI Number
1824206
Product Code
KNC
Status
Terminated
Root Cause
Other
Initiated
April 24, 2003
Posted
May 29, 2003
Terminated
September 25, 2003
Address
125 E Pearl St, Batesville, IN, 47006

Description

Hill-Rom brand Affinity three birthing bed

Reason

Possible electrical shock hazard, as the power cord grounding pin may break off or become detached.

Action

Letters dated April 24, 2003 were sent to each consignee instructing them to inspect the beds for power cords with non-flared ground pins, to cut plug off any such cords when found, and replace these plugs with new plugs with the flared ground pin. [The type of plug on the unit can be seen by the naked eye, as the plug is transparent plastic molding.]

Distribution

United States, Bahamas, Bermuda, Canada, Colombia, Costa Rica, Dominican Republic, Guam, Japan, Mexico, Panama, Saudi Arabia, South Korea, Taiwan and Venezuela