FDA Recall Terminated

Equinoxe Reverse Shoulder Fixed Angle Torque Defining Screw Kit

Recall: Z-0900-2017 · Initiated October 25, 2016

Recall

Recall Number
Z-0900-2017
Event Number
75740
Firm
Exactech, Inc.
FEI Number
1038671
Product Code
KWT
Status
Terminated
Root Cause
Component design/selection
Initiated
October 25, 2016
Terminated
July 26, 2019
Address
2320 NW 66th Ct, Gainesville, FL, 32653-1630

Description

Equinoxe Reverse Shoulder Fixed Angle Torque Defining Screw Kit

Reason

Disengaged screw head could potentially not be retained in the Torque Defining Screwdriver.

Action

Exactech sent an Advisory Notice on October 25, 2016, to Exactech field agents in the U.S. Agents outside of the U.S. were notified on November 2, 2016. Field agents were notified of the recall in the Advisory Notice and asked to notify their accounts that may have the recalled product in stock. Agents were required to complete and return the Field Advisory Notice Response Form within 5 working days. For further questions, please call (352) 377-1140.

Distribution

Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and to the states of : AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, MD, ME, MN, MO, NC, NJ, NV, NY, OH, OK, OR, PA, SC, TN, VA, WA and WI., and to the countries of : Austria, Australia, Canada, Germany, France, Spain, India, Switzerland, Italy, South Korea, United Kingdom, Brazil, Luxembourg, Greece, Guatemala, Netherlands, South Africa, and Sweden

Quantity

2,278 units