FDA Recall Terminated

Philips MX 16-Slice CT Scanner System, a whole-body computed tomography X-ray system.

Recall: Z-0899-2013 · Initiated December 19, 2012

Recall

Recall Number
Z-0899-2013
Event Number
64389
Firm
Philips Medical Systems (Cleveland) Inc
FEI Number
1525965
Product Code
JAK
Status
Terminated
Root Cause
Component design/selection
Initiated
December 19, 2012
Posted
February 28, 2013
Terminated
December 3, 2013
Address
595 Miner Rd, Cleveland, OH, 44143-2131

Description

Philips MX 16-Slice CT Scanner System, a whole-body computed tomography X-ray system.

Reason

The four screws that secure the patient table top to the carrier became loose, which caused the table top to detach.

Action

Philips Healthcare sent an Urgent - Medical Device Correction letter dated December 14, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to have BioMedical Engineering personnel inspect the patient table top screws that secure it to the carrier. Customers were instructed that If it is found that the table top screws are improperly seated, they should stop using the system and contact their local Philips Field Service Engineer for repair. A Philips Field Service Engineer will schedule a time to inspect and repair all the affected systems. For further information customers were instructed to contact their Philips Healthcare Customer Care Solutions Center at 1-800-722-9377 and follow the recorded menu option to reach a Customer Solutions. For questions regarding this recall call 800-722-9377, ext 5.

Distribution

Worldwide Distribution - USA including AL, AR, CA, FL, GA, IL, IN, KY, MI, MS, NC, NJ, NV, NY, OH, PA, PR and VA.

Quantity

412 Units