FDA Recall Terminated

Precedence SPECT/CT System, Precedence 16 3/8, Upper Patient Pallet, A single photon nuclear medicine and image from an X-ray computed tomography system, Model Number: 4535 602 50851, Catalog Number: 882350, Product is manufactured by Philips Medical Systems (Cleveland), Inc. Cleveland, OH Intended use: A single photon nuclear medicine and image from an X-ray computed tomography system.

Recall: Z-0897-2011 · Initiated January 26, 2009

Recall

Recall Number
Z-0897-2011
Event Number
51274
Firm
Phillips Nuclear Medicine
FEI Number
1525965
Product Code
INQ
Status
Terminated
Root Cause
Process change control
Initiated
January 26, 2009
Posted
January 12, 2011
Terminated
January 13, 2011
Address
540 Alder Dr, Bldg 4, Milpitas, CA, 95035-7443

Description

Precedence SPECT/CT System, Precedence 16 3/8, Upper Patient Pallet, A single photon nuclear medicine and image from an X-ray computed tomography system, Model Number: 4535 602 50851, Catalog Number: 882350, Product is manufactured by Philips Medical Systems (Cleveland), Inc. Cleveland, OH Intended use: A single photon nuclear medicine and image from an X-ray computed tomography system.

Reason

A problem has been identified with some upper patient pallets on Precedence SPECT/CT System, which may reduce its load bearing capacity and may result in its cracking. If this were to happens, could pose a risk to patient.

Action

The firm, Philips Healthcare, sent an "Urgent - Field Safety Notice" dated January 26, 2009 to all consignees/customers. The notice described the product, problem and action to be taken by customers and by Philips. The customers were instructed to reference their system's System Operation and System information Manuals regarding loading and unloading of patient; monitoring the patient during a study ,and careful monitoring during unsupported (cantilevered) table translation. Philips has ordered replacement parts and they will have a Philip Healthcare Field Service Engineer contact and visit the customers site within two months to replace the affected upper patient pallet. If you need any further information or support concerning this issue, please contact your local Philips representative at: Customer Care Center (1-800-722-9377, option 5: Diagnostic Imaging, option 5: Radiation Therapy, option 3: and Oncology or your local Philips Healthcare office.

Distribution

International distribution only: Mexico and Saudi Arabia.

Quantity

2 units