FDA Recall Open, Classified

Medtronic Synchromed II Model 8637 supplied in 20mL and 40 mL reservoir size and a catheter. Product Usage: The implantable Model 8637 SynchroMed II programmable pump is part of an infusion system that stores and delivers a prescribed drug to a specific site. The implanted infusion system consists of a Model 8637 SynchroMed II pump, supplied in 20 ml and 40 ml reservoir sizes, and a catheter. In the United States, the SynchroMed II Pump is specifically approved (indicated) for drugs as referenced in Indications Drug Stability and Emergency Procedures Reference Manual (refer to attached labeling # MA08927A006).

Recall: Z-0896-2018 · Initiated August 9, 2017

Recall

Recall Number
Z-0896-2018
Event Number
78609
Firm
Medtronic Neuromodulation
FEI Number
2182207
Product Code
LKK
Status
Open, Classified
Root Cause
Device Design
Initiated
August 9, 2017
Address
7000 Central Ave Ne, Minneapolis, MN, 55432-3568

Description

Medtronic Synchromed II Model 8637 supplied in 20mL and 40 mL reservoir size and a catheter. Product Usage: The implantable Model 8637 SynchroMed II programmable pump is part of an infusion system that stores and delivers a prescribed drug to a specific site. The implanted infusion system consists of a Model 8637 SynchroMed II pump, supplied in 20 ml and 40 ml reservoir sizes, and a catheter. In the United States, the SynchroMed II Pump is specifically approved (indicated) for drugs as referenced in Indications Drug Stability and Emergency Procedures Reference Manual (refer to attached labeling # MA08927A006).

Reason

Medtronic implemented a final design change intended to address the motor stall due to corrosion and is recalling devices with the previous design that are not implanted.

Action

Medtronic sent an Urgent Medical Device Recall letter dated September 2017, to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete and return the enclosed Customers Confirmation form. For questions contact Technical Services at 1-888-854-0978.

Distribution

US Nationwide Distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OR, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY

Quantity

27,547 US and 13,025 OUS