FDA Recall Terminated

Advanced Neuromodulation Systems, Brio DBS System implantable neurostimulator, IPG or R-IPG, Model 6788 Implantable Pulse Generator.

Recall: Z-0894-2013 · Initiated March 28, 2012

Recall

Recall Number
Z-0894-2013
Event Number
61554
Firm
Advanced Neuromodulation Systems Inc.
FEI Number
1627487
Product Code
MHY
Status
Terminated
Root Cause
Device Design
Initiated
March 28, 2012
Posted
February 28, 2013
Terminated
August 15, 2013
Address
6901 Preston Rd, Plano, TX, 75024-2508

Description

Advanced Neuromodulation Systems, Brio DBS System implantable neurostimulator, IPG or R-IPG, Model 6788 Implantable Pulse Generator.

Reason

Unintended Fluid ingress into the device header has been shown.

Action

St. Jude Medical sent an Important Product Field Corrective Action Device Information letter dated March 28, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check their inventory for any unused product, isolate the affected product for return to St. Jude Medical. Their local representatiave would collect the unused inventory. Any unused product would be replaced with product once inventory becomes available. For patients with implanted product and no degradation of therapy, it is recommended dthat the devices not be unnecesssarily explanted. For patient with implanted product who experience diminished or loss of therapy customers were advised to follow the steps provided in the letter. Customers were asked to forward the notification to any institution in the event one or more patients or products could be potentially affected by this action. Customers wsere asked to maintain a record of thenotice along with the recommendations to ensure the effectiveness of the communication. For questions regarding this recall call 972-309-2154.

Distribution

Worldwide Distribution - 53 consignees out of the US.

Quantity

388 units