GE Healthcare CIC Pro Clinical Information Center, Central Station, Software Version 4.1. , GE Medical Systems Information Technologies, Inc., 8200 W. Tower Ave. Milwaukee, Wisconsin USA. (System hardware consists of a processing unit and a monitor)
Recall
- Recall Number
- Z-0891-2007
- Event Number
- 37309
- Firm
- General Electric Medical Systems Information Technology
- FEI Number
- 2124823
- Product Code
- DSI
- Status
- Terminated
- Root Cause
- Other
- Initiated
- January 24, 2007
- Posted
- June 6, 2007
- Terminated
- December 10, 2011
- Address
- 8200 W Tower Ave, Milwaukee, WI, 53223-3219
Description
GE Healthcare CIC Pro Clinical Information Center, Central Station, Software Version 4.1. , GE Medical Systems Information Technologies, Inc., 8200 W. Tower Ave. Milwaukee, Wisconsin USA. (System hardware consists of a processing unit and a monitor)
Alarm and Reboot problems: Two separate issues could occur with the CIC Pro system when used with Telemetry. Each issue occurs in software version v4.1 only; A) the system may experience a loss of audible alarms when operating system memory resources reach 170MB and/or B) the system may intentionally or unexpectedly reboot .
GE Healthcare sent a certified letter titled "Urgent Medical Device Correction" to customers on 3/13/07 that explained the short and long term solutions and that the company would be providing a software update that will rectify the issue specific to CIC Pro v4.1 software.
Nationwide
5,190 Units